Specifications
What Cardia analyzes, what it needs to do so and how it fits into your hospital.
Population and modality
- Indicated population
- Adults, 18 years and older
- Modality
- 2D transthoracic echocardiography at rest
- ECG
- Not required. In case of arrhythmia, the specialist can reposition the systolic and diastolic frames
- Currently out of scope
- Stress echo, contrast, 3D, transesophageal, handheld devices, pediatric and neonatal, fetal echo
Required views
- B-mode (required)
- A2C, A4C and PLAX. PSAX optional, extends the analysis. At least one complete cardiac cycle per view
- Doppler (optional)
- A4C for mitral and tricuspid; A5C for aortic. Color, PW and CW per valve
- Strain (optional)
- A2C and A4C, already part of the B-mode, plus A3C for 18-segment GLS
Report contents
- Chambers
- Left atrium, left ventricle, right atrium, right ventricle
- Valves
- Mitral, tricuspid, aortic
- B-mode
- Chambers, walls and systolic function. Values indexed to body surface area
- Doppler
- Regurgitation: jet area and vena contracta. Stenosis: peak and mean gradients, valve area and PHT. Diastolic function: E/A, E/e′, septal and lateral e′
- Strain
- LV global longitudinal strain and 18 segments (6 per view in A2C, A3C and A4C), with polar map, curves and tracking
- Reference ranges
- American, European or the center's own guidelines. Strain per the 2025 ASE/EACVI consensus
Quality control
- Insufficient views
- Cardia detects them, alerts the specialist and does not issue a report it cannot support. That study is not billed
- Alternatives
- Upload a new acquisition or discard the metrics that depend on that view
- Report status
- Inconclusive until validated by the specialist
Integration and compatibility
- Input format
- Standard DICOM, independent of manufacturer or model
- Tested manufacturers
- Siemens, Philips and General Electric, among others
- Light integration
- Secure study upload and report in the browser. No installation or connection to core systems
- Full integration
- Sync from PACS and results in the electronic health record or viewer, with traceability for every study
- Hardware
- None additional. No proprietary probe or equipment
- Time to result
- Under three minutes from receipt of the DICOM
Security and data
- Encryption
- In transit and at rest
- Original DICOM
- Never modified
- Use of images
- Only for the study they belong to. Not used to train models without an explicit agreement